S72303

GENESYS 40 Vis Spectrophotometer

Manufacturer: Thermo Scientific

The price for this product is unavailable. Please request a quote

Baseline Flatness

±0.005 A

Connections

Single USB-ADuplex USB-AExport data to network or PC via USB, Ethernet, or Wi-Fi USB adaptor, availablePrint via USB, Ethernet or Wi-Fi USB adaptor available

Detector Type

Dual Silicon Photodiodes

Dimensions (L x W x H)

35.5 x 38.5 x 19.5 cm

Spectral Bandwidth

5 nm

Drift

(At 500 nm after 1 hour warmup) <0.0010 A/hr.

Keypad

Touchscreen

Noise

(RMS at 500nm 60 consecutive measurements)≤0.00020A at 0A at 260 and 500nm≤0.00030A at 1A at 260 and 500nm≤0.00040A at 2A at 260 and 500nm

Photometric Accuracy

±0.002 A at 0.5 A, ±0.004 A at 1.0 A, ±0.008 A at 2.0 A

Photometric Range

−3A to +3.5A

Printer Included

Snap-on printer available

Stray Light

<0.05%T at 340 nm and 400 nm

Wavelength Accuracy

±0.5 nm

Wavelength Repeatability

< ±0.2 nm

Compartments

Accessible from top, front or sideAble to accommodate cells up to 100mm cellsAvailable accessory for test tubes up to 25 mm diameter and 150 mm heightRemovable, washable sample compartment liner with magnetic placement and hold-down

Description

GENESYS 40 Visible Spectrophotometer

Display

7 in. High Definition Color Touchscreen, Fixed, 800 x 1280 pixels

Light Source

Tungsten Halogen

Wavelength Range

325 to 1100 nm

Electrical Requirements

External AC to DC converter. Voltage and Frequency (Hz) selected automatically, 100/240 V, 50/60 Hz

Lamp Life

>1000 hrs.

Optical Design

Double Beam

Photometric Display

−3A to +5A

Photometric Repeatability

±0.001 A at 1 A measured at 1.0 A at 546 nm

Scan Speed

Automatic – up to 1800 nm/min.

Warranty

2 years standard with 1 year extension upon registration of the instrument with Thermo Fisher Scientific within 6 months of purchase

Wavelength Data Interval

0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm

Weight (Metric)

7.5 kg

Related Products

Img

Thermo Scientific

S72304

--

Img

Thermo Scientific

S72305

--

Img

Thermo Scientific

S72306

--

Img

Thermo Scientific™

14-385-354

--

Img

Thermo Scientific™

14-385-355

--

Img

Thermo Scientific

S72307

--

Img

Thermo Scientific™

14-385-355PM

--

Img

Thermo Scientific

S28454

--

Description

  • The GENESYS 50 is a dual beam UV-Vis spectrophotometer in a single-cell configuration that meets all of the specifications for critical quality control applications and precise chemical measurements
  • The GENESYS 40 Dual Beam Vis Spectrophotometer is a visible spectral region version in this instrumental class
  • Both  feature data resolution from 0.2 nm to 5 nm
  • Features: 7-inch high-resolution touchscreen user interface Local control for speed and reliability Networking capable (Wi-Fi ready) Large sample compartment with front access Designed for: Chemistry labs Routine measurements Moderate throughput requirements