ZWACID012

ROCare

Acid cleaning chemical removes scaling from RO membranes

Manufacturer: Milli-Q

Synonym(S): ROCare A

Select a Size

Pack Size SKU Availability Price
12 EA ZWACID012-12-EA In Stock ₹ 37,130.01

ZWACID012 - 12 EA

₹ 37,130.01

In Stock

Quantity

1

Base Price: ₹ 37,130.01

GST (18%): ₹ 6,683.402

Total Price: ₹ 43,813.412

feature

removes scaling from RO membranes

packaging

pkg of 12 packs

compatibility

for use with Elix® Advantagefor use with Milli-Q® CLXfor use with Milli-Q® HRfor use with Milli-Q® HXfor use with Milli-Q® IX 7003for use with Milli-Q® IX 7005for use with Milli-Q® IX 7010for use with Milli-Q® IX 7015

shipped in

ambient

Related Products

Img

Milli-Q

ZWBASE012

Base cleaning chemical for removing...

Img

Milli-Q

ZMQUVLP01

For use with Milli-Q® Advantage A10...

Img

Milli-Q

EDSPAK0A1

Production of endocrine disruptor-f...

Img

Milli-Q

STARTPK01

Pretreatment attachments...

Description

  • General description: ROCare A is an acidic cleaner that protects the water purification systems by effectively removing some or most mineral scale buildup on the surface of the reverse osmosis (RO) membranes. The acid cleaner is delivered in a pouch and the empty pouch is removed after use. It is to be used periodically when prompted by the system, or when the RO % rejection drops more than 5%, or when the RO permeate flow rate drops more than 10% due to mineral scaling. The ROCare A pouch cleans RO membranes supporting Milli-Q® HR/HX/CLX, Milli-Q® IX 7003/05/10/15, and Elix® Advantage water purification systems.
  • Application: Removes scaling from RO membranes.
  • Features and Benefits: Anti-scaling agentpH 2 in an aqueous solutionContains 12 packages
  • Legal Information: ELIX is a registered trademark of Merck KGaA, Darmstadt, Germany

SAFETY INFORMATION

Pictograms

GHS05,GHS07

Signal Word

Danger

Hazard Statements

H302 - H314

Precautionary Statements

P280 - P301 + P330 + P331 - P305 + P351 + P338 - P308 + P311

Hazard Classifications

Acute Tox. 4 Oral - Skin Corr. 1B

Storage Class Code

8B - Non-combustible, corrosive hazardous materials